Deep-brain stimulation

A Human base model with an anchored deep-brain stimulation overlay: an implanted pulse generator in a subclavicular pocket, an extension lead tunnelled up the neck, and an intracranial electrode reaching a basal-ganglia target.

All configurations

Generator power cell · Implanted pulse generator · str skeletal system
Deep-brain stimulation on BodyParts3D meshes
Checking anatomy assets…
Device geometry is Corpus-authored schematic, not vendor CAD.

Configuration pathway

Generator to target: the deep-brain stimulation circuit

Trace an implanted pulse generator through its lead and cranial entry to a deep brain target, then read what is actually approved, what came through a humanitarian exemption, and what the reports on file record.

Powers implanted pulse generator

A deep-brain stimulation system is registered as an implanted pulse generator placed in a chest pocket, an extension lead tunnelled under the skin, and an electrode passing through a cranial entry point to a deep brain target. Stimulation settings are changed from outside the body with a clinician programmer.

Measured by
Generator service interval: settings-and-chemistry-dependent years.
Boundary
This is component topology only. Corpus renders the chain schematically and carries no lead trajectory, contact spacing, or target coordinate for any system.
  • The overlay is placed on reviewed attachment anchors that resolve to whole atlas structures. Anchor precision is region-level or orientation-only, so the rendered position indicates a body region, not a surgical location.
  • No vendor stimulation parameters, control gains, programming values, or fitting procedures are published here. Those are device-specific and clinician-set.
device registration

Component chain of a deep-brain stimulation system

A deep-brain stimulation system is registered as an implanted pulse generator placed in a chest pocket, an extension lead tunnelled under the skin, and an electrode passing through a cranial entry point to a deep brain target. Stimulation settings are changed from outside the body with a clinician programmer.

This is component topology only. Corpus renders the chain schematically and carries no lead trajectory, contact spacing, or target coordinate for any system.

Public regulatory and safety record

What the FDA record actually says

Every number below is read out of the openFDA device databases at build time. A product code names a regulatory category, not a product, so each row covers every device cleared or approved in that category, from every manufacturer.

Stimulator, Intracerebral/Subcortical, Implanted GYZ

Route to market
Premarket approval
Premarket approvals on file
None on file
Recall records
0
Reports received
73 (death 2, injury 31, malfunction 13, other 26, unspecified 1)
MAUDE is passive surveillance. These counts are reports received, not events that occurred. They have no denominator, are subject to under- and over-reporting, and cannot establish incidence, prevalence, or that a device caused the outcome described.

Recall pinning coverage: No recalls are listed under this code.

What the reports are coded as

Device problems

  1. Explanted24
  2. Device remains implanted16
  3. Replace14
  4. Other (for use when an appropriate device code cannot be identified)7
  5. Fracture5
  6. Lead(s), fracture of5
  7. Lead(s), breakage of4
  8. Reaction4
  9. Crack3
  10. Device Operates Differently Than Expected3
  11. Difficult To Position3

8 code entries among the most frequent record that no specific code applied.

Patient problems

  1. Therapy/non-surgical treatment, additional24
  2. Surgical procedure17
  3. Treatment with medication(s)14
  4. Surgical procedure, repeated10
  5. Unspecified Infection10
  6. Hospitalization required9
  7. Other (for use when an appropriate patient code cannot be identified)8
  8. Surgical procedure, additional7
  9. Drainage6
  10. Edema6HP:0000969
  11. Electric Shock6
  12. Cognitive Changes5
The 12 most frequent device-problem and patient-problem codes. One report can carry several codes, so they do not sum to the report total. Codes marked uncoded record that no specific code applied. An HPO term is shown only where the FDA code is, word for word, an HPO label or exact synonym; that is a vocabulary match, not a clinical mapping.

Pinning coverage: 0 of 11 listed coded terms are pinned to a part. Not pinned to a part yet: Explanted, Device remains implanted, Replace, Other (for use when an appropriate device code cannot be identified), Fracture, Lead(s), fracture of, Lead(s), breakage of, Reaction, Crack, Device Operates Differently Than Expected, Difficult To Position. Uncoded entries: Unknown (for use when the device problem is not known).

Labeled device identifiers: 0

AccessGUDID lists no device identifier under this product code.

AccessGUDID lists a device identifier for every labeled version and package, including external processors, batteries, cables, and accessories filed under the product code. The total counts labeled identifiers, not distinct products and not devices implanted. MRI safety is the labeling status the manufacturer submitted, not a clearance for any scan.

Stimulator, Electrical, Implanted, For Parkinsonian Tremor MHY

Route to market
Premarket approval
Premarket approvals on file
P140009, P960009
Recall records
46
Reports received
53,536 (death 742, injury 25,692, malfunction 26,869, other 196, unspecified 37)
MAUDE is passive surveillance. These counts are reports received, not events that occurred. They have no denominator, are subject to under- and over-reporting, and cannot establish incidence, prevalence, or that a device caused the outcome described.

Recall pinning coverage: 0 of 15 listed recalls are pinned to a part. Not pinned yet: Z-0023-2018, Z-0121-2024, Z-0184-2009, Z-0185-2009, Z-0186-2009, Z-0187-2009, Z-0201-2022, Z-0206-2025, Z-0217-2024, Z-0279-2014, Z-0290-2007, Z-0294-2022, Z-0370-06, Z-0418-2022, Z-0459-2026.

What the reports are coded as

Device problems

  1. High impedance7,391
  2. Device Operates Differently Than Expected5,009
  3. Battery Problem3,888
  4. Failure to Deliver Energy3,060
  5. Insufficient Device Problem Information2,917
  6. Connection Problem2,915
  7. Charging Problem2,600
  8. Low impedance2,594
  9. Inappropriate/Inadequate Shock/Stimulation2,451
  10. Break2,439
  11. Material Integrity Problem2,386

11,349 code entries among the most frequent record that no specific code applied.

Patient problems

  1. Therapeutic Response, Decreased5,172
  2. Unspecified Infection4,599
  3. Shaking/Tremors3,447
  4. Therapeutic Effects, Unexpected3,145
  5. Complaint, Ill-Defined2,988
  6. Electric Shock2,429
  7. Ambulation or Postural Difficulties2,076
  8. Pain1,569HP:0012531
  9. Inadequate Pain Relief1,308

22,034 code entries among the most frequent record that no specific code applied.

The 12 most frequent device-problem and patient-problem codes. One report can carry several codes, so they do not sum to the report total. Codes marked uncoded record that no specific code applied. An HPO term is shown only where the FDA code is, word for word, an HPO label or exact synonym; that is a vocabulary match, not a clinical mapping.

Pinning coverage: 4 of 11 listed coded terms are pinned to a part. Not pinned to a part yet: Device Operates Differently Than Expected, Failure to Deliver Energy, Insufficient Device Problem Information, Low impedance, Inappropriate/Inadequate Shock/Stimulation, Break, Material Integrity Problem. Uncoded entries: Adverse Event Without Identified Device or Use Problem.

Labeled device identifiers: 338
MRI safety labeling
Labeling does not contain MRI Safety Information 209 · MR Conditional 103 · MR Unsafe 26
Labelers with the most identifiers
MEDTRONIC, INC. 208 · BOSTON SCIENTIFIC NEUROMODULATION CORPORATION 86 · ST. JUDE MEDICAL, INC. 34 · ADVANCED NEUROMODULATION SYSTEMS, INC. 9 · Thoratec Corporation 1
Most common device types (GMDN)
Deep brain electrical stimulation system 201 · Multiple active implantable device programmer 59 · Neurosurgical procedure kit, non-medicated, single-use 25 · Neural-tissue electrical stimulation lead adaptor 12 · Noninvasive device battery charger 12 · Implant battery charger 8 · Implantable spinal cord electrical stimulation system programmer 7 · Subcutaneous tunneller, single-use 5
AccessGUDID lists a device identifier for every labeled version and package, including external processors, batteries, cables, and accessories filed under the product code. The total counts labeled identifiers, not distinct products and not devices implanted. MRI safety is the labeling status the manufacturer submitted, not a clearance for any scan.

Stimulator, Electrical, Implanted, For Parkinsonian Symptoms NHL

Route to market
Premarket approval
Premarket approvals on file
P150031, P960009
Recall records
4
Reports received
1,837 (death 42, injury 1,611, malfunction 182, other 2)
MAUDE is passive surveillance. These counts are reports received, not events that occurred. They have no denominator, are subject to under- and over-reporting, and cannot establish incidence, prevalence, or that a device caused the outcome described.

Recall pinning coverage: 0 of 4 listed recalls are pinned to a part. Not pinned yet: Z-1890-2024, Z-2125-2020, Z-2340-2025, Z-2558-2025.

What the reports are coded as

Device problems

  1. Unexpected Therapeutic Results337
  2. High impedance154
  3. Migration139
  4. Charging Problem82
  5. Wireless Communication Problem64
  6. Battery Problem63
  7. Inappropriate/Inadequate Shock/Stimulation43
  8. Use of Device Problem37
  9. Intermittent Energy Output36
  10. Fracture32
  11. Premature Discharge of Battery31

1,042 code entries among the most frequent record that no specific code applied.

Patient problems

  1. Bacterial Infection197
  2. Unspecified Infection186
  3. Swelling/ Edema155
  4. Shaking/Tremors149
  5. Movement Disorder111HP:0100022
  6. Pain97HP:0012531
  7. Erosion90
  8. Undesired Nerve Stimulation73
  9. Inadequate Pain Relief70

678 code entries among the most frequent record that no specific code applied.

The 12 most frequent device-problem and patient-problem codes. One report can carry several codes, so they do not sum to the report total. Codes marked uncoded record that no specific code applied. An HPO term is shown only where the FDA code is, word for word, an HPO label or exact synonym; that is a vocabulary match, not a clinical mapping.

Pinning coverage: 0 of 11 listed coded terms are pinned to a part. Not pinned to a part yet: Unexpected Therapeutic Results, High impedance, Migration, Charging Problem, Wireless Communication Problem, Battery Problem, Inappropriate/Inadequate Shock/Stimulation, Use of Device Problem, Intermittent Energy Output, Fracture, Premature Discharge of Battery. Uncoded entries: Adverse Event Without Identified Device or Use Problem.

Labeled device identifiers: 310
MRI safety labeling
Labeling does not contain MRI Safety Information 202 · MR Conditional 95 · MR Unsafe 13
Labelers with the most identifiers
MEDTRONIC, INC. 204 · BOSTON SCIENTIFIC NEUROMODULATION CORPORATION 92 · ADVANCED NEUROMODULATION SYSTEMS, INC. 9 · ST. JUDE MEDICAL, INC. 4 · Thoratec Corporation 1
Most common device types (GMDN)
Deep brain electrical stimulation system 180 · Multiple active implantable device programmer 59 · Neurosurgical procedure kit, non-medicated, single-use 25 · Neural-tissue electrical stimulation lead adaptor 10 · Noninvasive device battery charger 10 · Implant battery charger 8 · Implantable spinal cord electrical stimulation system programmer 7 · Subcutaneous tunneller, single-use 5
AccessGUDID lists a device identifier for every labeled version and package, including external processors, batteries, cables, and accessories filed under the product code. The total counts labeled identifiers, not distinct products and not devices implanted. MRI safety is the labeling status the manufacturer submitted, not a clearance for any scan.

Deep Brain Stimulator For Obsessive Compulsive Disorder (Ocd) OLM

Route to market
Humanitarian device exemption
Premarket approvals on file
None; this category reached the market through a humanitarian exemption
Recall records
0
Reports received
247 (death 2, injury 100, malfunction 145)
MAUDE is passive surveillance. These counts are reports received, not events that occurred. They have no denominator, are subject to under- and over-reporting, and cannot establish incidence, prevalence, or that a device caused the outcome described.

Recall pinning coverage: No recalls are listed under this code.

What the reports are coded as

Device problems

  1. Battery Problem50
  2. Connection Problem39
  3. High impedance37
  4. Material Integrity Problem36
  5. Charging Problem33
  6. Failure to Deliver Energy25
  7. Improper or Incorrect Procedure or Method20
  8. Insufficient Device Problem Information15
  9. Communication or Transmission Problem14
  10. Therapeutic or Diagnostic Output Failure13
  11. Malposition of Device11

46 code entries among the most frequent record that no specific code applied.

Patient problems

  1. Cognitive Changes27
  2. Unspecified Infection20
  3. Anxiety14HP:0000739
  4. Depression13HP:0000716
  5. Electric Shock12
  6. Emotional Changes10
  7. Irritability6HP:0000737
  8. Therapeutic Response, Decreased6
  9. Complaint, Ill-Defined5

181 code entries among the most frequent record that no specific code applied.

The 12 most frequent device-problem and patient-problem codes. One report can carry several codes, so they do not sum to the report total. Codes marked uncoded record that no specific code applied. An HPO term is shown only where the FDA code is, word for word, an HPO label or exact synonym; that is a vocabulary match, not a clinical mapping.

Pinning coverage: 0 of 11 listed coded terms are pinned to a part. Not pinned to a part yet: Battery Problem, Connection Problem, High impedance, Material Integrity Problem, Charging Problem, Failure to Deliver Energy, Improper or Incorrect Procedure or Method, Insufficient Device Problem Information, Communication or Transmission Problem, Therapeutic or Diagnostic Output Failure, Malposition of Device. Uncoded entries: Adverse Event Without Identified Device or Use Problem.

Labeled device identifiers: 121
MRI safety labeling
Labeling does not contain MRI Safety Information 90 · MR Conditional 31
Labelers with the most identifiers
MEDTRONIC, INC. 121
Most common device types (GMDN)
Multiple active implantable device programmer 51 · Deep brain electrical stimulation system 50 · Neurosurgical procedure kit, non-medicated, single-use 7 · Neural-tissue electrical stimulation lead adaptor 6 · Trial spinal cord/peripheral nerve percutaneous analgesic electrical stimulation system pulse generator 5 · Active implantable device control magnet 1 · Sacral plexus percutaneous incontinence-control electrical stimulation system pulse generator 1
AccessGUDID lists a device identifier for every labeled version and package, including external processors, batteries, cables, and accessories filed under the product code. The total counts labeled identifiers, not distinct products and not devices implanted. MRI safety is the labeling status the manufacturer submitted, not a clearance for any scan.
  • Educational reference only. Not a medical device, not diagnostic, and never stimulation programming or lead localization.
  • Device geometry is Corpus-authored and generic. It is not any manufacturer's lead or generator, and contact spacing, lead diameter, and generator size differ between products.
  • Corpus renders the basal ganglia as a single mesh. The subthalamic nucleus and globus pallidus internus are not separate geometry, so the target marker cannot show which nucleus a therapy uses.
  • The rendered lead trajectory is a schematic path between reviewed anchors. Real trajectories are planned per person from imaging and are not straight lines between landmarks.
  • OSS-DBS and Lead-DBS informed the generic topology only. Their assets and code are reference-only under the source lock and are not redistributed here.
  • Stimulation parameters, symptom outcomes, and adverse effects depend on target, programming, and diagnosis and are carried as evidence records rather than as properties of this geometry.